After preparing the buffer solution, it should be placed in a glass bottle or a polyethylene bottle (alkaline pH buffer solutions such as pH 9.18, pH 10.01, pH 12.46, etc. should be placed in polyethylene bottles). The bottle cap should be tightly closed and stored at low temperature (5-10 ℃) in the refrigerator. It can generally be used for about six months. If turbidity, mold or precipitation are found, it cannot be continued to be used. When using, prepare several 50mL polyethylene small bottles, pour the buffer solution from the large bottles into the small bottles, and let them stand at ambient temperature for 1-2 hours until the temperature is balanced before use. After use, it should not be poured into the large bottle to avoid contamination. The buffer solution in the bottle can be used for 2-3 days under environmental conditions of>10 درجة . بشكل عام، يمكن استخدام محاليل pH 7.00 وpH 6.86 وpH 14.00 لفترات زمنية أطول. تكون محاليل pH 9.18 وpH 10.01 عرضة لتغيرات الأس الهيدروجيني بسبب امتصاص ثاني أكسيد الكربون من الهواء.
ما هو الغرض من محلول الرقم الهيدروجيني المنظم؟
قم بمعايرة مقياس الأس الهيدروجيني قبل قياس الأس الهيدروجيني.
للتحقق من دقة مقياس الأس الهيدروجيني، على سبيل المثال، بعد معايرة مقياس الأس الهيدروجيني بـ 6.86 وpH 14.00، أدخل قطب الأس الهيدروجيني في محلول الأس الهيدروجيني 9.18 وتحقق مما إذا كانت قيمة عرض الجهاز متوافقة مع قيمة الأس الهيدروجيني للمحلول القياسي.
هل من الضروري إعادة معايرة مقياس الأس الهيدروجيني أثناء قياس الدقة العامة؟ بعد معايرة مقياس الأس الهيدروجيني واستخدامه، قد يحدث انجراف أو تغيرات. ولذلك، قبل الاختبار، أدخل القطب في محلول عازل قياسي قريب من المحلول المقاس، وحدد ما إذا كانت إعادة المعايرة ضرورية بناءً على حجم الخطأ.